SO2, NOx, O2, CO, CO2

Measuring principle
  • The light source emits ultraviolet beam which transfers to gas cell through optical fiber
  • After absorption by measuring gas, the light beam transfers through optical fiber to spectrometer.
  • Being dispersed through the grating, and converted into electric signal by the array sensor, a continuous absorption spectrum of measuring gas is thereafter obtained.

Technology Comparison
Item UV Hot-Wet   Method NDIR
Spectral   Range Complete and   continuous absorption spectra obtained through holographic grating and diode   array detection Non Dispersive,   characteristic absorption spectra obtained through band-pass filter
Linear   Response High wavelength   resolution secures linear response band-pass   filtration compromises linear response 
SO2 Loss Hot-wet   prohibits SO2   dissolution Condensation   inevitably causes SO2   loss
Sampling   System Heat tracing.   Requests only dust filtration, simple and no corrosive Cold-dry.   Requests dust filtration, condensation, water removal and corrosion proof
Zero   Calibration Period Long (DOAS   algorithm) Short
Anti-Interference The combination   of complete & continuous absorption spectra, high wavelength resolution,   DAOS algorithm erases interference from dust, moisture and background gases Vulnerable to   moisture interference
Gas Cell Connected by optical fiber,   easy to replace and low maintenance cost Complex to replace and   maintenance
Maintenance
Reliability No moving parts Moving parts   existed

OMA-2000

Features and Benefits
  • Full spectrum measurement and DOAS (Differential Optical Absorption Spectroscopy) algorithm, excelling in accuracy, repeatability and anti-interference with moisture, dust.
  • Modular design and no moving parts, higher reliability and easier maintenance.
  • Pulse Xenon Lamp used as light source with over 10-year lifetime.
  • The adoption of diode array sensor enables instant spectrum acquisition and rapid response.

Technical Data
SO2, NOx Analyzer OMA-2000
Technical   Principle UV-DOAS
Installation   type Hot-wet    Extractive
Measurement   Range 0~200ppm,    0~1000ppm, 0~3000ppm
Zero   Drift ≤    ±2%F.S.
Span   Drift ≤    ±2%F.S.
Accuracy ≤    ±1%
Response   Time(T90) ≤    60s
Linearity ≤    ±1%F.S.
O2  Analyzer ZrO-100/KE-25F3
Technical   Principle Zirconia/Electrochemical
Measurement   Range 0~25%
Response   Time(T90) ≤    60s
Linearity ≤    ±2%F.S.
CO, CO2  Analyzer TOC(CO,  CO 2)
Technical   Principle NDIR
Installation   type Extractive
CO   Measurement Range 0~1000    ppm
CO2   Measurement Range 0~20%
Zero   Drift ≤    ±2%F.S.
Span   Drift ≤    ±2%F.S.
Accuracy ≤    ±1%
Response   Time(T90) ≤    60s
Linearity ≤    ±2%F.S.